Abstract
PURPOSE: Poorly characterized macromolecular (MM) and baseline artefacts are known to reduce metabolite quantitation accuracy in (1) H MR spectroscopic imaging (MRSI). Increasing echo time (TE) and improvements in MM analysis schemes have both been proposed as strategies to improve metabolite measurement reliability. In this study, the influence of TE and two MM analysis schemes on MRSI reproducibility are investigated.
METHODS: An experimentally acquired baseline was collected using an inversion recovery sequence (TI = 750 ms) and incorporated into the analysis method. Intrasubject reproducibility of MRSI scans, acquired at 3 Tesla, was assessed using metabolite coefficients of variance (COVs) for both experimentally acquired and simulated MM analysis schemes. In addition, the reproducibility of TE = 35 ms, 80 ms, and 144 ms was evaluated.
RESULTS: TE = 80 ms was the most reproducible for singlet metabolites with COVs < 6% for total N-acetyl-aspartate, total creatine, and total choline; however, moderate multiplet dephasing was observed. Analysis incorporating the experimental baseline achieved higher Glu and Glx reproducibility at TE = 35 ms, and showed improvements over the simulated baseline, with higher efficacy for poorer data.
CONCLUSION: Overall, TE = 80 ms yielded the most reproducible singlet metabolite estimates. However, combined use of a short TE sequence and the experimental baseline may be preferred as a compromise between accuracy, multiplet dephasing, and T2 bias on metabolite estimates.
Original language | English |
---|---|
Pages (from-to) | 34–43 |
Journal | Magnetic Resonance in Medicine |
Volume | 77 |
Issue number | 1 |
Early online date | 22 Jan 2016 |
DOIs | |
Publication status | Published - 27 Dec 2016 |
Keywords
- macromolecule
- MRSI
- reproducibility
- TE
- baseline