Measurement of Frontal Midline Theta Oscillations Using OPM-MEG

Natalie Rhodes, Molly Rea, Elena Boto, Lukas Rier, Vishal Shah, Ryan M. Hill, James Osborne, Cody Doyle, Niall Holmes, Sebastian C. Coleman, Karen Mullinger, Richard Bowtell, Matthew J. Brookes*

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

38 Downloads (Pure)

Abstract

Optically pumped magnetometers (OPMs) are an emerging lightweight and compact sensor that can measure magnetic fields generated by the human brain. OPMs enable construction of wearable magnetoencephalography (MEG) systems, which offer advantages over conventional instrumentation. However, when trying to measure signals at low frequency, higher levels of inherent sensor noise, magnetic interference and movement artefact introduce a significant challenge. Accurate characterisation of low frequency brain signals is important for neuroscientific, clinical, and paediatric MEG applications and consequently, demonstrating the viability of OPMs in this area is critical. Here, we undertake measurement of theta band (4–8 Hz) neural oscillations and contrast a newly developed 174 channel triaxial wearable OPM-MEG system with conventional (cryogenic-MEG) instrumentation. Our results show that visual steady state responses at 4 Hz, 6 Hz and 8 Hz can be recorded using OPM-MEG with a signal-to-noise ratio (SNR) that is not significantly different to conventional MEG. Moreover, we measure frontal midline theta oscillations during a 2-back working memory task, again demonstrating comparable SNR for both systems. We show that individual differences in both the amplitude and spatial signature of induced frontal-midline theta responses are maintained across systems. Finally, we show that our OPM-MEG results could not have been achieved without a triaxial sensor array, or the use of postprocessing techniques. Our results demonstrate the viability of OPMs for characterising theta oscillations and add weight to the argument that OPMs can replace cryogenic sensors as the fundamental building block of MEG systems.
Original languageEnglish
Article number120024
Number of pages12
JournalNeuroImage
Volume271
Early online date12 Mar 2023
DOIs
Publication statusPublished - 1 May 2023

Keywords

  • Magnetoencephalography
  • Optically pumped magnetometers
  • Working memory
  • Neural oscillations
  • Theta oscillations
  • Low frequency

Fingerprint

Dive into the research topics of 'Measurement of Frontal Midline Theta Oscillations Using OPM-MEG'. Together they form a unique fingerprint.

Cite this