Strength of poly-ether-ether-ketone : effects of sterilisation and thermal ageing

Research output: Contribution to journalArticlepeer-review

11 Citations (Scopus)
1433 Downloads (Pure)

Abstract

This paper investigates the strength of polyether-ether-ketone (PEEK) after sterilisation and thermal ageing. PEEK specimens were divided into five groups, according to whether the specimens had been annealed, sterilised or aged. Specimens were subjected to either static or dynamic three-point bend tests. Static tests involved loading the specimens until a maximum displacement of 40 mm was reached. Dynamic tests involved applying a sinusoidally varying force at a frequency of 5 Hz. The maximum force applied to a specimen was based on a percentage of the static yield strength. Testing continued until failure or run out of 10 million cycles. Sterilisation and ageing resulted in no significant change in the static yield strength. Annealing was found to significantly increase the yield strength. For the dynamic tests, the fatigue strength was in the range 99.4 to 107.4 MPa; sterilisation and thermal ageing were found to have no effect on fatigue strength.
Original languageEnglish
Pages (from-to)1001-1005
Number of pages5
JournalPolymer Testing
Volume32
Issue number6
Early online date3 Jun 2013
DOIs
Publication statusPublished - 1 Sept 2013

Keywords

  • Fatigue strength;
  • Flexural strength;
  • Gamma sterilisation;
  • Poly-ether-ether-ketone (PEEK);
  • Thermal ageing