Metabolic profiling of serum using Ultra Performance Liquid Chromatography and the LTQ-Orbitrap mass spectrometry system

Warwick B Dunn, David Broadhurst, Marie Brown, Philip N Baker, Christopher W G Redman, Louise C Kenny, Douglas B Kell

Research output: Contribution to journalArticlepeer-review

158 Citations (Scopus)

Abstract

Advances in analytical instrumentation can provide significant advantages to the volume and quality of biological knowledge acquired in metabolomic investigations. The interfacing of sub-2 microm liquid chromatography (UPLC ACQUITY) and LTQ-Orbitrap mass spectrometry systems provides many theoretical advantages. The applicability of the interfaced systems was investigated using a simple 11-component metabolite mix and a complex mammalian biofluid, serum. Metabolites were detected in the metabolite mix with signals that were linear with their concentration over 2.5-3.5 orders of magnitude, with correlation coefficients greater than 0.993 and limits of detection less than 1 micromol L(-1). Reproducibility of retention time (RSD
Original languageEnglish
Pages (from-to)288-98
Number of pages11
JournalJournal of Chromatography. B, Analytical Technologies in the Biomedical and Life Sciences
Volume871
Issue number2
DOIs
Publication statusPublished - 15 Aug 2008

Fingerprint

Dive into the research topics of 'Metabolic profiling of serum using Ultra Performance Liquid Chromatography and the LTQ-Orbitrap mass spectrometry system'. Together they form a unique fingerprint.

Cite this