Characteristics of transcutaneous carbon dioxide tension monitors in normal adults and critically ill patients

S V Rithalia, T H Clutton-Brock, J Tinker

    Research output: Contribution to journalArticlepeer-review

    11 Citations (Scopus)

    Abstract

    The performance of three electrodes used for transcutaneous carbon dioxide (tcPCO2) monitoring is compared in 15 healthy volunteers and 26 critically ill adults. All three electrodes showed good correlation between tcPCO2 and arterial blood PCO2 (PaCO2) with a correlation coefficient (r) greater than 0.86. There was little difference in the performance characteristics of the three monitors. They may be usefully employed to estimate PaCO2 values when used with a modified calibration recommended by the manufacturers.