Abstract
The aim of this study was to test mechanically a new flexible fixation system for the lumbar spine. This device incorporates loops of polyester braid which are secured by a crimped titanium sleeve. In tensile tests, all loops failed, by slippage through the crimped sleeve, at 434±25 N (mean±standard deviation from five loops) for a single crimp and 415±15 N (from five loops) for two crimps. The intact system was then tested according to the ASTM standard. In a static test, all five specimens failed by slippage of the braid through the sleeve. Initial slippage occurred between 600 and 700 N, but the mean maximum load sustained was 1090±140 N. Dynamic tests were performed on ten constructs at a frequency of 5 Hz, under a range of loading conditions. The maximum load applied in any of the tests was 825 N. Two constructs did not complete the required 5×106 test cycles because of fracture of their spinous process hooks. However, other tests, under the same conditions, showed no signs of failure. Fracture occurred as a result of fretting damage from the recommended stainless steel roll pins.