Abstract
Opioid induced respiratory depression is potentially fatal. The aim of this study was to validate a monitoring system that could be used to assess respiratory depression in postoperative patients. The hypercapnic ventilatory response was estimated non-invasively in 12 volunteers. In two steps, we tested a system which delivered carbon dioxide (CO(2)) challenges through a venturi mask, measuring changes in ventilation with an uncalibrated respiratory inductance plethysmograph (RIP). RIP and pneumotachograph measurements of ventilation, taken at the same time during a CO(2) challenge, were similar; group mean (SD), pneumotachograph 13.9 (3.5) l x min(-1) x kPa(-1), RIP 14.3 (2.9) l x min(-1) x kPa(-1). Bland-Altmann analysis showed the variation between these two methods was +/- 5 l x min(-1) x kPa(-1) (2 SD). Second, we confirmed that the venturi mask is suitable for delivering CO(2) challenges. Despite the variability in RIP measurements, a simple multimodal respiratory monitoring system could be developed that incorporates clinical observation and non-invasive measurement of the ventilatory response to CO(2).
| Original language | English |
|---|---|
| Pages (from-to) | 426-432 |
| Number of pages | 7 |
| Journal | Anaesthesia |
| Volume | 60 |
| DOIs | |
| Publication status | Published - 1 May 2005 |
Fingerprint
Dive into the research topics of 'Evaluation of a non-invasive method of assessing opoid induced respiratory depression'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver