Abstract
Purpose
To compare cardiovascular magnetic resonance-feature tracking (CMR-FT) with spatial modulation of magnetization (SPAMM) tagged imaging for the calculation of short and long axis Lagrangian strain measures in systole and diastole.
Materials and Methods
Healthy controls (n = 35) and patients with dilated cardiomyopathy (n = 10) were identified prospectively and underwent steady-state free precession (SSFP) cine imaging and SPAMM imaging using a gradient-echo sequence. A timed offline analysis of images acquired at identical horizontal long and short axis slice positions was performed using CMR-FT and dynamic tissue-tagging (CIMTag2D). Agreement between strain and strain rate (SR) values calculated using these two different methods was assessed using the Bland–Altman technique.
Results
Across all participants, there was good agreement between CMR-FT and CIMTag for calculation of peak systolic global circumferential strain (−22.7 ± 6.2% vs. −22.5 ± 6.9%, bias 0.2 ± 4.0%) and SR (−1.35 ± 0.42 1/s vs. −1.22 ± 0.42 1/s, bias 0.13 ± 0.33 1/s) and early diastolic global circumferential SR (1.21 ± 0.44 1/s vs. 1.07 ± 0.30 1/s, bias −0.14 ± 0.34 1/s) at the subendocardium. There was satisfactory agreement for derivation of peak systolic global longitudinal strain (−18.1 ± 5.0% vs. −16.7 ± 4.8%, bias 1.3 ± 3.8%) and SR (−1.04 ± 0.29 1/s vs. −0.95 ± 0.32 1/s, bias 0.09 ± 0.26 1/s). The weakest agreement was for early diastolic global longitudinal SR (1.10 ± 0.40 1/s vs. 0.67 ± 0.32 1/s, bias −0.42 ± 0.40 1/s), although the correlation remained significant (r = 0.42, P < 0.01). CMR-FT generated these data over four times quicker than CIMTag.
Conclusion
There is sufficient agreement between systolic and diastolic strain measures calculated using CMR-FT and myocardial tagging for CMR-FT to be considered as a potentially feasible and rapid alternative.
To compare cardiovascular magnetic resonance-feature tracking (CMR-FT) with spatial modulation of magnetization (SPAMM) tagged imaging for the calculation of short and long axis Lagrangian strain measures in systole and diastole.
Materials and Methods
Healthy controls (n = 35) and patients with dilated cardiomyopathy (n = 10) were identified prospectively and underwent steady-state free precession (SSFP) cine imaging and SPAMM imaging using a gradient-echo sequence. A timed offline analysis of images acquired at identical horizontal long and short axis slice positions was performed using CMR-FT and dynamic tissue-tagging (CIMTag2D). Agreement between strain and strain rate (SR) values calculated using these two different methods was assessed using the Bland–Altman technique.
Results
Across all participants, there was good agreement between CMR-FT and CIMTag for calculation of peak systolic global circumferential strain (−22.7 ± 6.2% vs. −22.5 ± 6.9%, bias 0.2 ± 4.0%) and SR (−1.35 ± 0.42 1/s vs. −1.22 ± 0.42 1/s, bias 0.13 ± 0.33 1/s) and early diastolic global circumferential SR (1.21 ± 0.44 1/s vs. 1.07 ± 0.30 1/s, bias −0.14 ± 0.34 1/s) at the subendocardium. There was satisfactory agreement for derivation of peak systolic global longitudinal strain (−18.1 ± 5.0% vs. −16.7 ± 4.8%, bias 1.3 ± 3.8%) and SR (−1.04 ± 0.29 1/s vs. −0.95 ± 0.32 1/s, bias 0.09 ± 0.26 1/s). The weakest agreement was for early diastolic global longitudinal SR (1.10 ± 0.40 1/s vs. 0.67 ± 0.32 1/s, bias −0.42 ± 0.40 1/s), although the correlation remained significant (r = 0.42, P < 0.01). CMR-FT generated these data over four times quicker than CIMTag.
Conclusion
There is sufficient agreement between systolic and diastolic strain measures calculated using CMR-FT and myocardial tagging for CMR-FT to be considered as a potentially feasible and rapid alternative.
Original language | English |
---|---|
Pages (from-to) | 1000–1012 |
Journal | Journal of Magnetic Resonance Imaging |
Volume | 41 |
Issue number | 4 |
Early online date | 28 Mar 2014 |
DOIs | |
Publication status | Published - 19 Mar 2015 |
Keywords
- cine magnetic resonance imaging
- left ventricular function
- tagging
- feature tracking
Fingerprint
Dive into the research topics of 'Comparison of magnetic resonance feature tracking for systolic and diastolic strain and strain rate calculation with spatial modulation of magnetization imaging analysis'. Together they form a unique fingerprint.Projects
- 1 Finished
-
The Effects of a Modest Reduction in Renal Function on Cardiovascular Structure and Function: A Study of Kidney Donors
Townend, J. (Principal Investigator), Ferro, C. (Co-Investigator), Madhani, M. (Co-Investigator), Marshall, J. (Co-Investigator) & Moody, W. (Co-Investigator)
1/04/11 → 31/03/14
Project: Research