TY - JOUR
T1 - Cerebral oxygenation in traumatic brain injury
T2 - can a non-invasive frequency domain near-infrared spectroscopy device detect changes in brain tissue oxygen tension as well as the established invasive monitor?
AU - Davies, David
AU - Clancy, Michael
AU - Dehghani, Hamid
AU - Lucas, Samuel
AU - Forcione, Mario
AU - Yakoub, Kamal
AU - Belli, Antonio
PY - 2019/4/1
Y1 - 2019/4/1
N2 - The cost and highly invasive nature of brain monitoring modality in traumatic brain injury patients currently restrict its utility to specialist neurological intensive care settings. We aim to test the abilities of a frequency domain near-infrared spectroscopy (FD-NIRS) device in predicting changes in invasively measured brain tissue oxygen tension. Individuals admitted to a United Kingdom specialist major trauma center were contemporaneously monitored with an FD-NIRS device and invasively measured brain tissue oxygen tension probe. Area under the curve receiver operating characteristic (AUROC) statistical analysis was utilized to assess the predictive power of FD-NIRS in detecting both moderate and severe hypoxia (20 and 10 mm Hg, respectively) as measured invasively. Sixteen individuals were prospectively recruited to the investigation. Severe hypoxic episodes were detected in nine of these individuals, with the NIRS demonstrating a broad range of predictive abilities (AUROC 0.68-0.88) from relatively poor to good. Moderate hypoxic episodes were detected in seven individuals with similar predictive performance (AUROC 0.576-0.905). A variable performance in the predictive powers of this FD-NIRS device to detect changes in brain tissue oxygen was demonstrated. Consequently, this enhanced NIRS technology has not demonstrated sufficient ability to replace the established invasive measurement.
AB - The cost and highly invasive nature of brain monitoring modality in traumatic brain injury patients currently restrict its utility to specialist neurological intensive care settings. We aim to test the abilities of a frequency domain near-infrared spectroscopy (FD-NIRS) device in predicting changes in invasively measured brain tissue oxygen tension. Individuals admitted to a United Kingdom specialist major trauma center were contemporaneously monitored with an FD-NIRS device and invasively measured brain tissue oxygen tension probe. Area under the curve receiver operating characteristic (AUROC) statistical analysis was utilized to assess the predictive power of FD-NIRS in detecting both moderate and severe hypoxia (20 and 10 mm Hg, respectively) as measured invasively. Sixteen individuals were prospectively recruited to the investigation. Severe hypoxic episodes were detected in nine of these individuals, with the NIRS demonstrating a broad range of predictive abilities (AUROC 0.68-0.88) from relatively poor to good. Moderate hypoxic episodes were detected in seven individuals with similar predictive performance (AUROC 0.576-0.905). A variable performance in the predictive powers of this FD-NIRS device to detect changes in brain tissue oxygen was demonstrated. Consequently, this enhanced NIRS technology has not demonstrated sufficient ability to replace the established invasive measurement.
KW - brain tissue oxygen tension
KW - cerebral hypoxia
KW - cerebral non-invasive monitoring
KW - critical care
KW - frequency domain near-infrared spectroscopy
KW - traumatic brain injury
UR - http://www.scopus.com/inward/record.url?scp=85063270688&partnerID=8YFLogxK
U2 - 10.1089/neu.2018.5667
DO - 10.1089/neu.2018.5667
M3 - Article
C2 - 29877139
AN - SCOPUS:85063270688
SN - 0897-7151
VL - 36
SP - 1175
EP - 1183
JO - Journal of Neurotrauma
JF - Journal of Neurotrauma
IS - 7
ER -