Abstract
Background: Rapid expansion of antiretroviral therapy in Malawi has occurred in the relative absence of suitable pediatric CD4 counting facilities. We have recently validated in adults a simplified affordable flow cytometric CD4 counting method, the Blantyre count. There is a need for this technology to transfer to government laboratories run by local staff, and to be validated in children, where %CD4/lymphocyte values are required. Methods: We assessed agreement of %CD4/lymphocyte values determined by the Blantyre count and Panleucogate methods on an EPICS XL-MCL flow cytometer on 113 venous blood samples from HIV-seropositive children in Blantyre, Malawi. All assays were performed by two Malawian laboratory technicians. Results: Overall bias between the two methods was -0.13% (95% CI -0.37 to 0.11) and limits of agreement were -2.69 to 2.43% (95% CI -3.11 to -2.27 and 2.01 to 2.85). Limits of agreement were within -3.00 and 3.00 for each laboratory technician. Coefficient of variation for the Blantyre count assay was 2.0% and samples showed good stability over 5 days. Conclusions: The Blantyre count method can accurately determine %CD4/lymphocyte values in blood of HIV-seropositive children on an EPIC XL-MCL flow cytometer at a reagent cost of US $0.21 per test or less. The assay can be competently carried out by local laboratory technicians. (C) 2008 Clinical Cytometry Society.
Original language | English |
---|---|
Pages (from-to) | S90-S97 |
Journal | Cytometry Part B: Clinical Cytometry |
Volume | 74B |
Issue number | S1 |
Early online date | 1 Jan 2008 |
DOIs | |
Publication status | Published - 1 Jan 2008 |